Yokogawa EJX440A User Manual
Page 71
A-5
IM 01C25T01-01E
Checklist (2/5)
No.
Items
Parameters
Result
Example
8
Alarm status
• Check the alarm status shown in
Diag Error.
• Check that the alarm status of “ILBD over
range” is not shown in
Diag Error.
Diag Error
A Blocking
Large Fluct L
Large Fluct H
L Side Blocking
H Side Blocking
B Blocking
Invalid Ref BlkF
Invalid Ref SPH
Invalid Ref SPL
Invalid Ref DP
ILBD over range
9
ILBD parameters
• Record the values of parameters for ILBD
operation.
• Check the status of parameters for ILBD
operation.
*: Record the value after checked that the
status of each parameter is “GOOD”.
Diag Period
180
Lim fDPmax
3.000000
Lim fDPmin
0.300000
Lim fSPlmax
5.000000
Lim fSPlmin
0.500000
Lim fSPhmax
5.000000
Lim fSPhmin
0.500000
Lim BlkFmax
0.600000
Lim BlkFmin
-0.600000
Lim DPavgmax
1.000000
Lim DPAvgmin
0.050000
Diag Supp Count
3
Ref fDP*
7.43245E-09
Ref fSPl*
7.25765E-09
Ref fSPh*
7.18374E-09
Ref DPAvg*
5.36425E+00
fDP*
7.48562E-09
fSPl*
7.23277E-09
fSPh*
7.14085E-09
BlkF*
-0.287259
DPAvg*
0.055957