1) setting password, Operation – Yokogawa EJX440A User Manual
Page 35
<3. Operation>
3-25
IM 01C25T01-01E
(1) Setting Password
F0341.ai
(ENTER)
(DEL)
EJX:
Hot key
1 keypad input
2 Wrt protect menu
HELP
SEND
ESC
ENTER
Press
Hot key.
Select
Wrt protect menu.
Set
1 2 3 4
and press
ENTER (F4).
EJX:
Enter new password to
change state of write
protect:
1234
HELP
DEL
ABORT ENTER
(ENTER)
EJX:
Re-enter new password
withing 30 seconds:
1234
1234
HELP
DEL
ABORT ENTER
Example: Set the password to
1 2 3 4 .
Ч4
Ч2
Set the new password again, and
press
ENTER (F4).
EJX:
Wrt protect menu
1 Write Protect No
2 Enable wrt 10min
3 New password
4 Software seal Keep
HELP
SEND
ESC
ENTER
Select
New password.
2
3
4
1
(OK)
Press
OK (F4).
Write Protect status changes
from
NO to YES.
EJX:
It changed the state
of protection related
password.
HELP
DEL
ABORT
OK
5
(2) Entering Password to Enable the Parameter
Changes
F0342.ai
(ENTER)
(DEL)
EJX:
Hot key
1 keypad input
2 Wrt protect menu
HELP
SEND
ESC
ENTER
Press
Hot key.
Select
Wrt protect menu.
Set
1 2 3 4
and press
ENTER (F4).
EJX:
Enter current
password to enable to
write for 10
minutes:
1234
HELP
DEL
ABORT ENTER
(OK)
EJX:
Released the write
protection for 10
minutes.
HELP
DEL
ABORT
OK
(OK)
Press
OK (F4), and the write
protect is released for 10 minutes.
EJX:
If you wish to
release completely,
you have to change
password to all of
spaces.
HELP
DEL
ABORT
OK
Example: Enter the password of
1 2 3 4 .
×4
Press
ENTER (F4).
EJX:
Wrt protect menu
1 Write Protect No
2 Enable wrt 10min
3 New password
4 Software seal Keep
HELP
SEND
ESC
ENTER
Select
Enable wrt 10min.
2
3
4
5
1
(3) Releasing Password
To release the password completely, enter spaces
in the
New password field while the write protect
function is released.
(4) Software Seal
When you lose the password that has been
registered, it is possible to release the
Write
Protect mode by using the general use password:
“YOKOGAWA.” When the password is used, the
status shown in the parameter of
Software seal
is changed from “KEEP” to “BREAK.” The status
returns to “KEEP” by entering a newly set password
at
Enable wrt 10min.