Yokogawa EJX440A User Manual
Page 63
<5. Parameter Summary>
5-2
IM 01C25T01-01E
Function
Label
Item
Contents
Default value
Handling
Fast key
sequences
Diag Lim
Lim fDPmax
Upper limit for Ratio fDP
As specified
model
W, G
1, 2, 4, 3, 2, 3, 1
Lim fDPmin
Lower limit for Ratio fDP
As specified
model
W, G
1, 2, 4, 3, 2, 3, 2
Lim fSPlmax
Upper limit for Ratio fSPl
As specified
model
W, G
1, 2, 4, 3, 2, 3, 3
Lim fSPlmin
Lower limit for Ratio fSPl
As specified
model
W, G
1, 2, 4, 3, 2, 3, 4
Lim fSPhmax
Upper limit for Ratio fSPh
As specified
model
W, G
1, 2, 4, 3, 2, 3, 5
Lim fSPhmin
Lower limit for Ratio fSPh
As specified
model
W, G
1, 2, 4, 3, 2, 3, 6
Lim BlkFmax
Upper limit for BlkF
As specified
model
W, G
1, 2, 4, 3, 2, 3, 7
Lim BlkFmin
Lower limit for BlkF
As specified
model
W, G
1, 2, 4, 3, 2, 3, 8
Lim DPAvgmax
Upper limit for DPAvg
As specified
model
W, G
1, 2, 4, 3, 2, 3, 9
Lim DPAvgmin
Lower limit for DPAvg
As specified
model
W, G
1, 2, 4, 3, 2, 3, a
Diag Mode
Diag Mode
ILBD operation mode
Stop, Calculation, or
Reference
W, G
1, 2, 4, 3, 1, 1
Diag Option
Diag Option
Alarm masking
W, G
1, 2, 4, 2
Diag Output
Diag Out Option
Output mode of 4-20mA when
an advanced diagnostic alarm is
generated
Off, Burnout, or Fall back
Off
W, G
1, 2, 4, 5, 1
Diag Fixed Out Val 4-20 mA output when an advanced
diagnostic alarm is generated
3.6000 to 21.6000 mA
21.6 mA
W, G
1, 2, 4, 5, 2
Diag Period Diag Period
Sampling period per one Diag count
180 sec
W, G
1, 2, 4, 3, 2, 2
Diag
Reference
Diag Description
Memo field
32 alphanumerics
W, G
1, 2, 4, 3, 2, 4, 1
Ref fDP
Reference value of fDP
W, G
1, 2, 4, 3, 2, 4, 2
Ref fDP Status
Status of Reference fDP
R, G
1, 2, 4, 3, 1, 4, 2
Ref fSPl
Reference value of fSPl
W, G
1, 2, 4, 3, 2, 4, 3
Ref fSPl Status
Status of Reference fSPl
R, G
1, 2, 4, 3, 1, 4, 4
Ref fSPh
Reference value of fSPh
W, G
1, 2, 4, 3, 2, 4, 4
Ref fSPh Status
Status of Reference fSPh
R, G
1, 2, 4, 3, 1, 4, 6
Ref BlkF
Reference value of BlkF
W, G
1, 2, 4, 3, 2, 4, 5
Ref BlkF Status
Status of Reference BlkF
R, G
1, 2, 4, 3, 1, 4, 8
Ref DPAvg
Reference value of DPAvg
W, G
1, 2, 4, 3, 2, 4, 6
Ref DPAvg Status
Status of Reference DPAvg
R, G
1, 2, 4, 3, 1, 4, ↓
Diag Ref Lim Ref Lim fDPmin
Lower limit of fDP
7.0E-10
W, G
1, 2, 4, 3, 2, 6, 1
Ref Lim fSPmin
Lower limit of fSPl and fSPh
1.0E-10
W, G
1, 2, 4, 3, 2, 6, 2
Ref Lim BlkFmax
Upper limit of BlkF
0.5
W, G
1, 2, 4, 3, 2, 6, 3
Diag Supp
Count
Diag Supp Count
Detection count to generate an alarm
W, G
1, 2, 4, 3, 2, 5
Diag
Variables
Ratio fDP
SQRT (fDP/Ref fDP).
R, G
1, 2, 4, 3, 1, 3, 2
Ratio fDP Status
Status of Ratio fDP
R, G
1, 2, 4, 3, 1, 3, 1
Ratio fSPl
SQRT (fSPl/Ref fSPl).
R, G
1, 2, 4, 3, 1, 3, 4
Ratio fSPl Status
Status of Ratio fSPl
R, G
1, 2, 4, 3, 1, 3, 3
Ratio fSPh
SQRT (fSPl/Ref fSPl).
R, G
1, 2, 4, 3, 1, 3, 6
Ratio fSPh Status
Status of Ratio fSPh
R, G
1, 2, 4, 3, 1, 3, 5
BlkF
Blockage degree characterized in
comparison of high-pressure side
and low-pressure side pressure
fluctuation values
R, G
1, 2, 4, 3, 1, 3, 8
BlkF Status
Status of BlkF
R, G
1, 2, 4, 3, 1, 3, 7
DPAvg
Ratio of the average of differential
pressure/pressure to the maximum
span of an EJX transmitter.
R, G
1, 2, 4, 3, 1, 3, ↓
DPAvg Status
Status of DPAvg
R, G
1, 2, 4, 3, 1, 3, 9
CRatio fDP
Compnesated fDP
R, G
1, 2, 4, 3, 1, 3, ↓
CRatio fDP
Status of Cratio fDP
R, G
1, 2, 4, 3, 1, 3, ↓
NRatio fDP
Non-compnesated fDP
R, G
1, 2, 4, 3, 1, 3, ↓
NRatio fDP Status
Status of Nratio fDP
R, G
1, 2, 4, 3, 1, 3, ↓
*1:
Handling:
R=Read only, W=Read & Write, M=Method, A=Applicable for option code AL, G=Applicable for option code DG6,
D=Applicable for differential pressure transmitters. Do not change these parameters for pressure transmitters.
*2:
The default value shows MWP (Maximum working pressure) of the capsule.
Since the working pressure limit varies according to the Model, refer to the General Specifications section in each user’s manual.