Aux Output
The following signals can be output through the AUX OUTPUT coaxial
connector on the back of the WaveSurfer MXs-B oscilloscope (this coaxial
connector is not labeled).
Aux Output Off - turns off the auxiliary output signal
Trigger Out - can be used to trigger another oscilloscope
Trigger Enabled - can be used as a gating function to trigger
another instrument when the oscilloscope is ready
Pass/Fail - allows pulse duration setup from 1 ms to 500
ms. Generates a pulse when pass/fail testing is active and
conditions are met.
Setting Up Auxiliary Output
1. Touch Utilities Utilities Setup... from the menu bar.
2. Touch the Aux Output tab.
3. Touch one of the buttons under Use Auxiliary Output For.
4. If you selected Pass/Fail, touch inside the Pulse Duration field and
enter a value from 1 ms to 500 ms, using the pop-up numeric keypad.
NOTE: The CAL output on the front of the WaveSurfer MXs-B oscilloscope is
always set to a 1V / 1 kHz square wave.