Documentation, Configuration – Ronan X96S MOLD LEVEL GAGE User Manual
Page 45

41
Documentation
For future calibrations, document these items:
(a) Environmental/process conditions that influence the low calibrate / high calibrate. The next time a calibration is performed,
you will need to duplicate the conditions, or account for the differences.
(b) All changes made to factory-default settings such as filtering values, calibrations values, detector counts, etc.
(c) The information from the status displays. A record of "counts" being received from the detector may assist with future
troubleshooting efforts.
Configuration
Ronan ships the Mold Level Monitor System with factory-default software settings. Those settings are responsible for the
information that initially appears on the status displays.
After installation at your site, you may need to reconfigure the system to fit your application. The goal is to correlate the X96S
output with your actual level readings. The list below summarizes the activities that are detailed in the remainder of this chapter:
- Check the factory-default settings to be sure they are appropriate for your circumstances. IF NOT, make the necessary
changes and document those changes for future reference.
- Perform an initial calibration to correlate the X96S’s output to the actual mold level
- Document the detector counts during the low calibration and the high calibration. Document the actual level during the low
calibration and the high calibration. Note if the mold water was running and if the oscillator was running. Also, record
changes you make to factory-default settings. Keep this information for future reference.