Dio output, Relay and ttl menus, Output – Ronan X96S MOLD LEVEL GAGE User Manual
Page 24: Cont’d.)
20
DIO Output
This menu is used to view and configure the DIO output. It contains the following items:
ITEM
FUNCTION
Output
Shows and allows user to select and configure a specific digital output (Relay 1-4
or TTL 1-4)
Select Source
Selecting this allows user to assign an array of sources to the above digital output
Polarity
Shows and allows user to set the above digital output
Polarity has the following options to assign:
ITEM
FUNCTION
NO/Not Driven
Allows user to configure the selected output as non-fail safe mode
NC/Driven
Allows user to configure the selected output as a fail safe mode
Open/Not Driven
Allows user to force the selected output open or not driven (relay de-energized) or
driven (TTL not driven) regardless of the state of the source
Closed/Driven
Allows user to force the selected output closed (relay energized) or driven (TTL
driven) regardless of the state of the source
Relay and TTL Menus
The Relay and TTL menus are used to configure the X96S Relay Outputs and the 4 TTL Outputs. The Relay and TTL menus
show the settings of the corresponding 4 Relay Outputs and 4 TTL Outputs, allowing the characteristics of the outputs to be
changed. Each menu contains the following items:
Output
(Cont’d.)
ITEM
FUNCTION
Relay 1
Selecting this item takes the user to the Relay 1 menu
Relay 2
Selecting this item takes the user to the Relay 2 menu
Relay 3
Selecting this item takes the user to the Relay 3 menu
Relay 4
Selecting this item takes the user to the Relay 4 menu
TTL 1
Selecting this item takes the user to the TTL 1 menu
TTL 2
Selecting this item takes the user to the TTL 2 menu
TTL 3
Selecting this item takes the user to the TTL 3 menu
TTL 4
Selecting this item takes the user to the TTL 4 menu
Select Source and Polarity have the same options as Relays 1 through 4.