Step 4 High Calibration
(Insert a test block into the mold covering the full range of
measurement)
Scroll down
↓ to “Calibration” Press the Enter Key →
Scroll down
↓ to “Calibrate” Press the Enter Key →
Scroll down
↓ to “High Calibrate” Press the Enter Key →
Select “Perform High Cal” press the Enter Key
→
Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reaches 0.
Acknowledge the High Counts= XXXX by pressing the F4 key.
Scroll down
↓ to “High Cal Level” Press the Enter Key →
Notice the value that is displayed.
Use
← Key and → Key to toggle back and forth between digits.
Change the value using
↑ or ↓
Enter the value for the Full level (Example 6.0000 inches.).
Press The F4 Key to enter and store the value.
Press F3 Key (Home) to get back to the Main Menu.
Press F3 Key (Lock) to get back to the Status Display.
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Main Menu
Variables
Status Display
Configuration
Digital Outputs
Digital Inputs
Auto Cal
→Calibration
.
Diagnostics
LOCK
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
→Perform High Cal
.
Calibr. Mode Full M
High Cal Level 0.000
High Cal Cnts 126
HOME
C
F1
F2
F3
F4
Ronan Engineering
X96S
Mold Level
Calibration
Ref Constants
→Calibrate
.
Last Ref.Date 03/22
Last Ref.Time 0912
Loop Config
MLM Outputs Cfg
HOME
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Calibrate
Low Calibrate
→High Calibrate
.
HOME
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
Calibrating gage...
10 seconds left
ABORT
OK
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
High Counts = 132
Press OK to accept
or ABORT to cancel
calibration
ABORT
OK
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
WARNING – Remove
loop from automatic
control and the mold
is full or source
shutter is closed
ABORT
OK
F1
F2
F3
F4
Ronan Engineering
X96S
Mold Level
High Calibtate
Perform High Cal
Calibr. Mode Full M
→High Cal Level 6.00
High Cal Cnts 132
HOME
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Status Display
Level 6.0000 in
MLM Ch.1 6.0000 in
Head Temp 92.00 degF
Filt Counts 132
Raw Counts 131
Time 08/04/06 08:05
Diagnostics