37

Step 4 High Calibration

(Insert a test block into the mold covering the full range of
measurement)
Scroll down

↓ to “Calibration” Press the Enter Key

Scroll down

↓ to “Calibrate” Press the Enter Key

Scroll down

↓ to “High Calibrate” Press the Enter Key

Select “Perform High Cal” press the Enter Key

Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reaches 0.
Acknowledge the High Counts= XXXX by pressing the F4 key.

Scroll down

to “High Cal Level” Press the Enter Key

Notice the value that is displayed.
Use

Key and Key to toggle back and forth between digits.

Change the value using

↑ or

Enter the value for the Full level (Example 6.0000 inches.).
Press The F4 Key to enter and store the value.

Press F3 Key (Home) to get back to the Main Menu.
Press F3 Key (Lock) to get back to the Status Display.

C

F1

F2

F3

F4

Ronan Engineering

X96S Mold Level
Main Menu

Variables
Status Display
Configuration

Digital Outputs
Digital Inputs

Auto Cal

Calibration

.

Diagnostics

LOCK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate

Perform High Cal

.

Calibr. Mode Full M
High Cal Level 0.000
High Cal Cnts 126

HOME

C

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Calibration

Ref Constants

Calibrate

.

Last Ref.Date 03/22

Last Ref.Time 0912
Loop Config

MLM Outputs Cfg

HOME

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Calibrate
Low Calibrate

High Calibrate

.

HOME

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate

Calibrating gage...

10 seconds left

ABORT

OK

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate
High Counts = 132

Press OK to accept
or ABORT to cancel
calibration

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate
WARNING – Remove
loop from automatic
control and the mold
is full or source
shutter is closed

ABORT

OK

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

High Calibtate
Perform High Cal
Calibr. Mode Full M

High Cal Level 6.00

High Cal Cnts 132

HOME

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 6.0000 in

MLM Ch.1 6.0000 in
Head Temp 92.00 degF
Filt Counts 132
Raw Counts 131

Time 08/04/06 08:05

Diagnostics

Lo-C Hi-C Back