40

Step 1

Low Calibration using the
Remote Auto Cal Feature

(Have the mold empty with mold water and oscillator running )

Locate the Remote Low Auto-Cal pushbutton and the Auto-Calibration
indication. Hold the pushbutton for 3 to 5 seconds until the Auto-
Calibration indication turns on. Continued to run the mold in the empty
condition until the Auto-Reference indication turns off.

X96S Mold Level output should indicate zero (example: 4.00 mA or 0
Vdc). If not, repeat steps 1 & 2 again.

Auto-Cal

Push Button

Auto-Cal

Light

Step 2

High Calibration using the
Remote Auto Cal Feature

(Insert the test block into them mold covering the full range of
the measurement
)

Locate the High Auto-Calibrate pushbutton and the Auto-Calibrate
indication. Hold pushbutton for 3 to 5 seconds until the Auto-Calibrate
indication turns on. Continued to hold the test-block in front of the mold
until the Auto-Calibrate indication turns off.

X96S Mold Level output should indicate full (example: 20.00 mA or 10.00
Vdc). If not repeat steps 1 & 2 again.