Ronan X96S MOLD LEVEL GAGE User Manual
Page 44
40
Step 1
Low Calibration using the
Remote Auto Cal Feature
(Have the mold empty with mold water and oscillator running )
Locate the Remote Low Auto-Cal pushbutton and the Auto-Calibration
indication. Hold the pushbutton for 3 to 5 seconds until the Auto-
Calibration indication turns on. Continued to run the mold in the empty
condition until the Auto-Reference indication turns off.
X96S Mold Level output should indicate zero (example: 4.00 mA or 0
Vdc). If not, repeat steps 1 & 2 again.
Auto-Cal
Push Button
Auto-Cal
Light
Step 2
High Calibration using the
Remote Auto Cal Feature
(Insert the test block into them mold covering the full range of
the measurement )
Locate the High Auto-Calibrate pushbutton and the Auto-Calibrate
indication. Hold pushbutton for 3 to 5 seconds until the Auto-Calibrate
indication turns on. Continued to hold the test-block in front of the mold
until the Auto-Calibrate indication turns off.
X96S Mold Level output should indicate full (example: 20.00 mA or 10.00
Vdc). If not repeat steps 1 & 2 again.