38

Step 1

Low Calibration using the
Local Auto Cal Feature

(Have the mold empty with mold water and oscillator running )

Press F1 Key (Lo-C) to perform an Auto Low calibration.

Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reach 0.

Acknowledge the Low Counts = XXXX by pressing the F4 Key.

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 4.7500 in

MLM Ch.1 4.7500 in
Head Temp 92.00 degF
Filt Counts 1489
Raw Counts 1502
Time 08/04/06 08:05
Diagnostics

Lo-C Hi-C Back

C

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Status Display

WARNING –

Remove

loop from automatic
control, verify that
source is on and the
mold is empty

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Status Display

Calibrating gage...

10 seconds left

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Status Display

Low Counts = 18742

Press OK to accept

or ABORT to cancel

calibration

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 0.0000 in

MLM Ch.1 0.0000 in
Head Temp 92.00 degF
Filt Counts 18742
Raw Counts 18751

Time 08/04/06 08:05

Diagnostics

Lo-C Hi-C Back