Ronan X96S MOLD LEVEL GAGE User Manual
Page 43
39
Step 2
High Calibration using the
Local Auto-Cal Feature
(Insert a test block into the mold covering the full range of
measurement)
Press F2 Key (Hi-C) to perform an Auto High calibration.
Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reaches 0.
Acknowledge the High Counts = XXXX by pressing the F4 Key
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Status Display
Level 4.7500 in
MLM Ch.1 4.7500 in
Head Temp 92.00 degF
Filt Counts 1489
Raw Counts 1502
Time 08/04/06 08:05
Diagnostics
Lo-C Hi-C Back
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
WARNING – Remove
loop from automatic
control and the mold
is full or source
shutter is closed
ABORT
OK
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Status Display
Calibrating gage...
10 seconds left
ABORT
OK
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
High Calibrate
High Counts = 132
Press OK to accept
or ABORT to cancel
calibration
ABORT
OK
C
F1
F2
F3
F4
Ronan Engineering
X96S Mold Level
Status Display
Level 6.0000 in
MLM Ch.1 6.0000 in
Head Temp 92.00 degF
Filt Counts 132
Raw Counts 131
Time 08/04/06 08:05
Diagnostics
Lo-C Hi-C Back