Ronan X96S MOLD LEVEL GAGE User Manual

Page 43

background image

39

Step 2

High Calibration using the
Local Auto-Cal Feature

(Insert a test block into the mold covering the full range of
measurement)

Press F2 Key (Hi-C) to perform an Auto High calibration.

Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reaches 0.

Acknowledge the High Counts = XXXX by pressing the F4 Key

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 4.7500 in

MLM Ch.1 4.7500 in
Head Temp 92.00 degF
Filt Counts 1489
Raw Counts 1502
Time 08/04/06 08:05
Diagnostics

Lo-C Hi-C Back

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate
WARNING – Remove
loop from automatic
control and the mold
is full or source
shutter is closed

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display
Calibrating gage...

10 seconds left

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
High Calibrate
High Counts = 132

Press OK to accept
or ABORT to cancel
calibration

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 6.0000 in

MLM Ch.1 6.0000 in
Head Temp 92.00 degF
Filt Counts 132
Raw Counts 131

Time 08/04/06 08:05

Diagnostics

Lo-C Hi-C Back