beautypg.com

Step 3 low calibration – Ronan X96S MOLD LEVEL GAGE User Manual

Page 40

background image

36

C

F1

F2

F3

F4

Ronan Engineering

X96S Mold Level
Main Menu

Variables
Status Display
Configuration

Digital Outputs
Digital Inputs

Auto Cal

Calibration

.

Diagnostics

LOCK

Step 3 Low Calibration

(Have the mold empty with mold water and oscillator running )

Scroll down

↓ to “Calibration” Press the Enter Key

Scroll down

↓ to “Calibrate” Press the Enter Key

Scroll down

↓ to “Low Calibrate” Press the Enter Key

Select “Perform Low Cal” press the Enter Key

Read the next screen carefully and acknowledge the screen by
pressing the F4 Key. Wait until you see the counter reach 0.
Acknowledge the Low Counts = XXXX by pressing the F4 Key.

Scroll down

to “Low Cal Level” Press the Enter Key

Notice the value that is displayed.
Use

Key and Key to toggle back and forth between digits.

Change the value using

↑ or

Enter the value of the process (Example 0.0000 inches for Empty
Mold).
Press the F4 Key to enter and store the value.

Press F3 Key (Home) to get back to the Main Menu.
Press F3 Key (Lock) to get back to the Status Display.

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Low Calibtate

Perform Low Cal

.

Calibr. Mode Empty
Low Cal Level 0.000
Low Cal Cnts 18426

HOME

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Calibration

Ref Constants

Calibrate

.

Last Ref.Date 03/22
Last Ref.Time 0912
Loop Config
MLM Outputs Cfg

HOME

C

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Calibrate

Low Calibrate

High Calibrate

HOME

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Low Calibrate

Calibrating gage...

10 seconds left

ABORT

OK

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Low Calibrate

Low Counts = 18742

Press OK to accept

or ABORT to cancel

calibration

ABORT

OK

C

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Low Calibrate

WARNING –

Remove

loop from automatic
control, verify that
source is on and the
mold is empty

ABORT

OK

F1

F2

F3

F4

Ronan Engineering
X96S

Mold Level

Low Calibtate
Perform Low Cal
Calibr. Mode Empty

Low Cal Level 0.0000

Low Cal Cnts 18742

HOME

F1

F2

F3

F4

Ronan Engineering
X96S Mold Level
Status Display

Level 0.0000 in

MLM Ch.1 0.0000 in
Head Temp 92.00 degF
Filt Counts 18742
Raw Counts 18751

Time 08/04/06 08:05

Diagnostics

Lo-C Hi-C Back