Calibrate Menu

This menu is used to access the various calibration procedures. It contains the following items:

ITEM

FUNCTION

Low Calibrate

Selecting this item takes the user to the Low Calibrate menu

High Calibrate

Selecting this item takes the user to the High Calibrate menu

Low Calibrate Menu

This menu is used to perform the low calibrate procedure. It contains the following items:

ITEM

FUNCTION

Perform Low Cal

This item invokes a method that performs the low calibrate procedure

Calibr. Mode

This allows the user to select the method of calibration (empty mold/process/absorber)

Low Cal Level

Shows and allows the user to set the low cal level value

Low Cal Counts

Shows the raw captured counts for the low calibration

High Calibrate Menu

This menu is used to perform the high calibrate procedure. It contains the following items:

ITEM

FUNCTION

Perform High Cal

This item invokes a method that performs the high calibrate procedure

Calibr. Mode

This allows the user to select the method of calibration (empty mold/process/absorber)

High Cal Level

Shows and allows the user to set the high cal level value

High Cal Counts

Shows the raw captured counts for the high calibration

Loop Config Menu

This menu is used to access the primary 4-20mA loop calibration procedures. It contains the following items:

ITEM

FUNCTION

Loop test

This item invokes a method that performs a test on the primary 4-20mA current loop

Damping

Shows and allows the user to set the damping constant for the primary 4-20mA current
loop

D/A trim

This item invokes method that performs the D/A trimming of the primary 4-20mA
current loop

MLM Outputs Cfg Menu

This menu is used to access the MLM Outputs configuration. It contains the following items:

ITEM

FUNCTION

MLM Out 1 Test

This item invokes a method that performs a test on the primary 4-20mA (or 0-10 volt
MLM Out 1) loop

MLM Out 1 Trim

This item invokes a method that performs the D/A trimming of the MLM1 4-20mA (or
0 – 10 volts) loop

MLM Out 2 Test

This item invokes a method that performs a test on the primary 4-20mA (or 0-10 volt
MLM Out 2) loop

MLM Out 2 Trim

This item invokes a method that performs the D/A trimming of the MLM2 4-20mA (or
0 – 10 volts) loop