Electrical Installation of Interconnect Wiring

DO NOT APPLY POWER until wiring is carefully
checked.

Wire the equipment according to the detailed
interconnect drawing which is included in the Drawing
Chapter of this manual.

Follow local and national electrical codes for all
interconnections.

Consider the following guidelines before making any
electrical connections:

Use continuous conduit runs and protect housing junction
boxes from dripping of condensed moisture off of conduit.

Plug unused conduit holes to prevent entry of dirt and moisture.

Run the interconnect cable in a separate conduit. Feed the
cable through the conduit starting at the detector end and
terminate at the microprocessor end.

DO NOT run AC power cable in the same conduit with any of
the low-level cables (signal, mV, mA, etc.)

Maintain transient-free AC power sources between 105-130
VAC for the microprocessor. DO NOT use a line that is
connected to a large motor, welding equipment, solenoids, etc.

WITH POWER OFF - - -
Connect cable pre-wired MS connector to detector.

Immediately replace lid of detector housing to keep out
water and dirt.

Check connections at microprocessor chassis terminals.
Verify that all wires are fully inserted in terminal
sockets and the screws firmly tightened.