Calibration mode – Teledyne GFC-7001T - Trace CO Analyzer User Manual
Page 90
OverviewTeledyne API – Model T300/T300M CO AnalyzerTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
90
Figure 4-3:
Viewing and Clearing GFC7001T/GFC7001TM WARNING Messages
4.2. CALIBRATION MODE
Pressing the CAL button switches the GFC7001T/GFC7001TM into calibration mode.
In this mode the user can calibrate the instrument with the use of calibrated zero or span
gases. This mode is also used to check the current calibration status of the instrument.
For more information about setting up and performing standard calibration
operations or checks, see Section 9.
For more information about setting up and performing EPA equivalent calibrations,
see Section 10.
If the instrument includes one of the available zero/span valve options, the
SAMPLE mode display will also include CALZ and CALS buttons. Pressing
either of these buttons also puts the instrument into calibration mode.
The CALZ button is used to initiate a calibration of the analyzer’s zero point using
internally generated zero air.
The CALS button is used to calibrate the span point of the analyzer’s current
reporting range using span gas.
For more information concerning calibration valve options, see Table 1-1.
For information on using the automatic calibration feature (ACAL) in conjunction with
the one of the calibration valve options, see Section 9.4.
I
MPORTANT
I
MPACT ON
R
EADINGS OR
D
ATA
It is recommended that this span calibration be performed at 80-90% of
full scale of the analyzer’s currently selected reporting range.
EXAMPLES: If the reporting range is set for 0 to 50 ppm, an appropriate
span point would be 40-45 ppm. If the of the reporting range is set for 0
to 1000 ppb, an appropriate span point would be 800-900 ppb.