Setting the gfc7001t concentration alarm limits, Teledyne analytical instruments 137, Alarm status limit set point – Teledyne GFC-7001T - Trace CO Analyzer User Manual
Page 137
![background image](/manuals/345733/137/background.png)
SetupTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
137
a Single Pole Double Throw relay output accessible via the alarm output connector on
the instrument’s back panel (See Section 3.3.1.4). If the CO concentration measured by
the instrument rises above that limit, the alarm‘s status output relay is closed.
The default settings for ALM1 and ALM2 are:
Table 5-10:
CO Concentration Alarm Default Settings
ALARM STATUS
LIMIT
SET
POINT
1
alm1 Disabled
100
ppm
alm2 Disabled
300
ppm
1
Set points listed are for PPM. Should the reporting range units of measure be changed (See
Section 5.4.3) the analyzer will automatically scale the set points to match the new range unit
setting.
Note
To prevent the concentration alarms from activating during span
calibration operations ensure that the CAL or CALS button is pressed
prior to introducing span gas into the analyzer.
5.10.1. SETTING THE GFC7001T CONCENTRATION ALARM LIMITS
To enable either of the CO concentration alarms and set the limit points, press:
SETUP X.X
CO ALRM 2: OFF
ON
ENTR EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM VARS DIAG
ALRM
EXIT
DIAG FCAL CO ALARM 2=300.00 PPM
3
0
0
.0
0
ENTR
EXIT
Toggle these
buttons to Set
the alarm point
EXIT discards the new
setting
ENTR accepts the
new setting
SAMPLE
RANGE=50.0 PPM CO= XX.XX
TST> CAL
SETUP
SETUP X.X
CO ALRM 2, DISALBED
NEXT
EDIT
PRNT EXIT
Toggle this button to
enable/disable
the alarm
Continue pressing NEXT until the desired
Alarm is selected