SetupTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

133

5.9.3.11. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION

To configure the analyzer’s external analog inputs option, define for each channel:

 gain (number of units represented by 1 volt)
 offset

(volts)

 engineering units to be represented in volts (each press of the touchscreen button scrolls the

list of alphanumeric characters from A-Z and 0-9)

 whether to display the channel in the Test functions

To access and adjust settings for the external Analog Inputs option channels press:

DIAG AIO

XIN1 GAIN:1.00V/V

SET> EDIT

EXIT

DIAG

ANALOG I / O CONFIGURATION

PREV

NEXT

ENTR

EXIT

DIAG AIO

AOUTS CALIBRATED: NO

< SET SET> CAL

EXIT

DIAG AIO

XIN1 GAIN:1.00V/V

+ 0 0 1 .0 0 ENTR EXIT

Press to change

Gain value

DIAG AIO

XIN1:1.00,0.00,V,OFF

< SET SET> EDIT

EXIT

Press SET> to scroll to the first
channel. Continue pressing SET>
to view each of 8 channels.

Pressing ENTR records the new setting

and returns to the previous menu.

Pressing EXIT ignores the new setting and

returns to the previous menu.

Press EDIT at any channel
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).

DIAG AIO

XIN1 OFFSET:0.00V

< SET

SET> EDIT

EXIT

DIAG AIO

XIN1 UNITS:V

< SET SET> EDIT

EXIT

DIAG AIO

XIN1 DISPLAY:OFF

< SET EDIT

EXIT

Figure 5-7.

DIAG – Analog Inputs (Option) Configuration Menu

5.9.4. ELECTRICAL TEST

The electrical test function creates a current, which substitutes the PMT signal, and
feeds it into the preamplifier board. This signal is generated by circuitry on the pre-
amplifier board itself and tests the filtering and amplification functions of that assembly
along with the A/D converter on the motherboard. It does not test the PMT itself. (See
also Section

12.5.7.2).