5.9.3.11. ANALOG INPUTS (XIN1…XIN8) OPTION CONFIGURATION
To configure the analyzer’s external analog inputs option, define for each channel:
gain (number of units represented by 1 volt)
offset
(volts)
engineering units to be represented in volts (each press of the touchscreen button scrolls the
list of alphanumeric characters from A-Z and 0-9)
whether to display the channel in the Test functions
To access and adjust settings for the external Analog Inputs option channels press:
DIAG AIO
XIN1 GAIN:1.00V/V
SET> EDIT
EXIT
DIAG
ANALOG I / O CONFIGURATION
PREV
NEXT
ENTR
EXIT
DIAG AIO
AOUTS CALIBRATED: NO
< SET SET> CAL
EXIT
DIAG AIO
XIN1 GAIN:1.00V/V
+ 0 0 1 .0 0 ENTR EXIT
Press to change
Gain value
DIAG AIO
XIN1:1.00,0.00,V,OFF
< SET SET> EDIT
EXIT
Press SET> to scroll to the first
channel. Continue pressing SET>
to view each of 8 channels.
Pressing ENTR records the new setting
and returns to the previous menu.
Pressing EXIT ignores the new setting and
returns to the previous menu.
Press EDIT at any channel
to to change Gain, Offset,
Units and whether to display
the channel in the Test
functions (OFF/ON).
DIAG AIO
XIN1 OFFSET:0.00V
< SET
SET> EDIT
EXIT
DIAG AIO
XIN1 UNITS:V
< SET SET> EDIT
EXIT
DIAG AIO
XIN1 DISPLAY:OFF
< SET EDIT
EXIT
Figure 5-7.
DIAG – Analog Inputs (Option) Configuration Menu
5.9.4. ELECTRICAL TEST
The electrical test function creates a current, which substitutes the PMT signal, and
feeds it into the preamplifier board. This signal is generated by circuitry on the pre-
amplifier board itself and tests the filtering and amplification functions of that assembly
along with the A/D converter on the motherboard. It does not test the PMT itself. (See
also Section
12.5.7.2).