Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

259

12.1.4.3. RELAY BOARD STATUS LEDS

There are eight LEDs located on the Relay Board. The most important of which is D1,
which indicates the health of the I

2

C bus. If D1 is blinking the other faults following

LEDs can be used in conjunction with DIAG menu signal I/O to identify hardware
failures of the relays and switches on the relay (see Section 12.1.3 and Appendix D).

Table 12-4: I

2

C Status LED Failure Indications

LED

FUNCTION

FAULT STATUS

INDICATED FAILURE(S)

D1

(Red)

I

2

C bus Health

(Watch Dog

Circuit)

Continuously ON

or

Continuously OFF

Failed/Halted CPU
Faulty Motherboard, Touchscreen or Relay Board
Faulty Connectors/Wiring between Motherboard,
Touchscreen or Relay Board
Failed/Faulty +5 VDC Power Supply (PS1)

STATUS LED’s

DC VOLTAGE TEST

POINTS

RELAY PCA

PN 04135

Figure 12-5:

Relay Board Status LEDs