beautypg.com

Repair procedures, Repairing sample flow control assembly, Teledyne analytical instruments 281 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 281

background image

Troubleshooting and ServiceTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

281

12.6. REPAIR PROCEDURES

This contains procedures that might need to be performed on rare occasions when a
major component of the analyzer requires repair or replacement.

12.6.1. REPAIRING SAMPLE FLOW CONTROL ASSEMBLY

The critical flow orifice is housed in the flow control assembly (TAI P/N 001760400)
located on the top of the optical bench. A sintered filter protects the jewel orifice so it is
unusual for the orifice to need replacing, but if it does, or the filter needs replacement
please use the following procedure (see the Spare Parts list in Appendix B for part
numbers and kits):
1. Turn off power to the analyzer.
2. Locate the assembly. See Figure 3-6.
3. Disconnect the pneumatic connection from the flow assembly and the assembly

from the pump (if present)..

4. Remove the fitting and the components as shown in the exploded view below.
5. Replace the o-rings (P/N OR0000001) and the sintered filter (P/N FL0000001).
6. If replacing the critical flow orifice itself (P/N 000941000), make sure that the side

with the colored window (usually red) is facing upstream to the flow gas flow.

7. Apply new Teflon

®

tape to the male connector threads.

8. Re-assemble in reverse order.
9. After reconnecting the power and pneumatic lines, flow check the instrument as

described in Section 11.3.3.

Pneumatic Connector, Male 1/8”
(P/N FT_70

Spring
(P/N HW_20)

Sintered Filter
(P/N FL_01)

Critical Flow Orifice
(P/N 00094100)
Make sure it is placed with the
jewel down)

O-Ring

(P/N OR_01)

Purge Housing

(P/N 000850000)

Figure 12-14: Critical Flow Restrictor Assembly/Disassembly