beautypg.com

Teledyne analytical instruments 50 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 50

background image

Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

50

CONTROL IN

A B C D E F U

+

LO

W S

P

A

N

ZER

O

CONTROL IN

A B C D E F U

+

LOW

SP

AN

ZER

O

-

+

5 VDC Power

Supply

Local Power Connections

External Power Connections

HIGH SP

A

N

HIGH SP

AN

Figure 3-12:

Control Input Connector

Table 3-7: Control Input Signals

INPUT #

STATUS DEFINITION

ON CONDITION

A

REMOTE ZERO CAL

The analyzer is placed in Zero Calibration mode. The mode field of the
display will read ZERO CAL R.

B

REMOTE SPAN CAL

The analyzer is placed in span calibration mode as part of performing a low
span (midpoint) calibration. The mode field of the display will read LO CAL
R
.

C

REMOTE CAL HIGH RANGE

The analyzer is forced into high range for zero or span calibrations. This
only applies when the range mode is either DUAL or AUTO. The mode field
of the display will read HI CAL R.

D, E

& F

SPARE

Digital Ground

The ground level from the analyzer’s internal DC power supplies (same as
chassis ground).

U

External Power input

Input pin for +5 VDC required to activate pins A – F.

+

5 VDC output

Internally generated 5V DC power. To activate inputs A – F, place a jumper
between this pin and the “U” pin. The maximum amperage through this port
is 300 mA (combined with the analog output supply, if used).

3.3.1.7. CONCENTRATION ALARM RELAY (OPTION 61) STANDARD

CONFIGURATION

The concentration alarm option is comprised of four (4) “dry contact” relays on the rear
panel of the instrument. This relay option is different from and in addition to the
“Contact Closures” that come standard on all TAI instruments. Each relay has 3 pins:
Normally Open (NO), Common (C), and Normally Closed (NC).