Setup diag: diagnostics functions, Teledyne analytical instruments 112 – Teledyne GFC-7001T - Trace CO Analyzer User Manual
Page 112
SetupTeledyne API – Model T300/T300M CO Analyzer
Teledyne
Analytical
Instruments
112
To access and navigate the VARS menu, use the following button sequence.
SETUP X.X 4) CLOCK_ADJUST=0 Sec/Day
PREV
NEXT
JUMP
EDIT ENTR EXIT
SETUP X.X CLOCK_ADJUST=0 Sec/Day
+
0
0
ENTR EXIT
Enter sign and number of
seconds per day the clock
gains (-) or loses(+).
SETUP X.X
DYN_SPAN=OFF
OFF
ENTR EXIT
SETUP X.X
DYN_ZERO=OFF
OFF
ENTR EXIT
Toggle this button to turn
the Dynamic Zero
calibration feature ON/
OFF.
Concentration display
continuously cycles
through all gasses.
SAMPLE
RANGE=500.0 PPM CO= XXXX
TST> CAL
SETUP
SETUP X.X
0) DAS_HOLD_OFF=15.0 Minutes
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
SETUP X.X
SECONDARY SETUP MENU
COMM
VARS
DIAG
EXIT
SETUP X.X
ENTER PASSWORD:818
8
1
8
ENTR
EXIT
Toggle these
buttons to enter
the correct
PASSWORD.
In all cases:
EXIT discards the new
setting.
ENTR accepts the
new setting.
SETUP X.X
2) DYN_ZERO=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
3) DYN_SPAN=OFF
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
DAS_HOLD_OFF=15.0 Minutes
1
5
.0
ENTR EXIT
Toggle these buttons to set
the iDAS HOLDOFF time
period in minutes
(MAX = 20 minutes).
Toggle this button to turn
the Dynamic Span
calibration feature ON/
OFF.
SETUP X.X
1) CONC_PRECISION=AUTO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
CONC_PRECISION=AUTO
AUTO
1
2
3
4
ENTR EXIT
SETUP X.X
5) STABIL_GAS=CO
PREV
NEXT
JUMP
EDIT
PRNT EXIT
SETUP X.X
STABIL_GAS=O2
CO
CO2
O2
ENTR EXIT
Use these buttons to select
which gas will be reported
by the sTABIL test
function.
(O
2
is only available if the
optional O
2
sensor is
installed)
Use these buttons to select
the precision of the o33
concentration display.
Press NEXT for additional
VARS; press NEXT or
PREV to move back and
forth throughout the list of
VARS.
5.9. SETUP DIAG: DIAGNOSTICS FUNCTIONS
A series of diagnostic tools is grouped together under the SETUPMOREDIAG
menu, as these parameters are dependent on firmware revision (see Appendix A). These
tools can be used in a variety of troubleshooting and diagnostic procedures and are