Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

194

9.1.1.1. ZERO AIR

Zero air or zero calibration gas is defined as a gas that is similar in chemical
composition to the measured medium but without the gas to be measured by the
analyzer.
For the GFC7001T/GFC7001TM zero air should contain less than 25 ppb of CO and
other major interfering gases such as CO and Water Vapor. It should have a dew point
of -5

C or less.

If your application is not a measurement in ambient air, the zero calibration gas should
be matched to the composition of the gas being measured.
 Pure nitrogen (N

2

) can be used as a zero gas for applications where CO is measured

in nitrogen.

 If your analyzer is equipped with an external zero air scrubber option, it is capable

of creating zero air from ambient air.

For analyzers without the zero air scrubber, a zero air generator such as the Teledyne
T701 can be used. Please visit the company website for more information.

9.1.1.2. SPAN GAS

Span Gas is a gas specifically mixed to match the chemical composition of the type of
gas being measured at near full scale of the desired measurement range. It is
recommended that the span gas used have a concentration equal to 80-90% of the full
measurement range.
If Span Gas is sourced directly from a calibrated, pressurized tank, the gas mixture
should be CO mixed with Zero Air or N

2

at the required ratio.

For oxygen measurements using the optional O

2

sensor, we recommend a reference gas

of 21% O

2

in N

2

.

 For quick checks, ambient air can be used at an assumed concentration of 20.8%.
 Generally, O

2

concentration in dry, ambient air varies by less than 1%.

9.1.1.3. CALIBRATION GAS STANDARDS AND TRACEABILITY

All equipment used to produce calibration gases should be verified against standards of
the National Institute for Standards and Technology (NIST). To ensure NIST
traceability, we recommend to acquire cylinders of working gas that are certified to be
traceable to NIST Standard Reference Materials (SRM). These are available from a
variety of commercial sources.

Table 9-1:

NIST-SRMs Available for Traceability of CO Calibration Gases

NIST-SRM Type Nominal

Concentration

680b

CO in N

2

500

ppm

1681b

CO in N

2

1000

ppm

2613a

CO in Zero Air

20 ppm

2614a

CO in Zero Air

45 ppm

2659a1 O

2

in N

2

21% by weight

2626a CO

2

in N

2

4% by weight

27452 CO

2

in N

2

16% by weight

1

Used to calibrate optional O

2

sensor.

2

Used to calibrate optional CO

2

sensor.