beautypg.com

Teledyne analytical instruments 30 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 30

background image

SpecificationsTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

30

Parameter

Specification

Optional I/O

1 USB com port
1 RS485
8 analog inputs (0-10V, 12-bit)
4 digital alarm outputs (2 opto-isolated and 2 dry contact)
Multidrop RS232
3 4-20mA current outputs

Temperature Range

5 - 40

C operating, 10 - 40C EPA Equivalency (GFC7001T only)

Humidity Range

0-95% RH, Non-Condensing

Temp Coefficient

< 0.05 % per

C (minimum 50 ppb/C)

Voltage Coefficient

< 0.05 % per V

Dimensions (HxWxD)

7" x 17" x 23.5" (178 mm x 432 mm x 597 mm)

Weight

GFC7001T: 40 lbs (18.1 kg); GFC7001TM: 38.4 lbs (17.2)

Environmental Conditions

Installation Category (Over voltage Category) II Pollution Degree 2

1

As defined by the USEPA

2

At constant temperature and pressure

Table 2-2:

O

2

Sensor Option Specifications

Parameter Description

Ranges

0-1% to 0-100% user selectable. Dual ranges and auto-ranging supported.

Zero Noise

1

<0.02% O

2

Lower Detectable Limit

2

<0.04%

O

2

Zero Drift (24 hours)

3

<± 0.02% O

2

Zero Drift (7 days)

<±- 0.05% O

2

Span Noise

1

<

±

0.05% O

2

Span Drift (7 days)

<± 0.1% O

2

Accuracy

(intrinsic error) <± 0.1% O

2

Linearity

<± 0.1 % O

2

Temp Coefficient

<± 0.05% O

2

/°C,

Rise and Fall Time

<60 seconds to 95%

1

As defined by the USEPA

2

Defined as twice the zero noise level by the USEPA

3

Note: zero drift is typically <± 0.1% O

2

during the first 24 hrs of operation

Table 2-3:

CO

2

Sensor Option Specifications

Parameter Description

Ranges

0-1% to 0-20% user selectable. Dual ranges and auto-ranging supported.

Zero Noise

1

<0.02% CO

2

Lower Detectable Limit

2

<0.04%

CO

2

Zero Drift (24 hours)

<± 0.02% CO

2

Zero Drift (7 days)

<± 0.05% CO

2

Span Noise

1

<± 0.1% CO

2

Span Drift (7 days)

<± 0.1% CO

2