This section presents specifications for the GFC7001T/GFC7001TM analyzer and for its
second gas sensor options, EPA equivalency designation, and compliance statements.
2.1. SPECIFICATIONS
Table 2-1: GFC7001T/GFC7001TM Basic Unit Specifications
Parameter
Specification
Ranges
Min: 0-1 ppm Full scale
Max: 0-1,000 ppm Full scale (selectable, dual ranges and auto ranging supported)
Measurement Units
GFC7001T: ppb, ppm, µg/m
3
, mg/m
3
(user selectable)
GFC7001TM: ppm, mg/m
3
(user selectable)
Zero Noise
1
GFC7001T: < 0.02 ppm RMS
GFC7001TM: ≤ 0.1 ppm RMS
Span Noise
1
GFC7001T:<0.5% of rdg RMS over 5ppm
3
GFC7001TM:>0.5% of rdg RMS over 20ppm
Lower Detectable Limit
1
GFC7001T: < 0.04 ppm
GFC7001TM: 0.2 ppm
Zero Drift (24 hours)
2
GFC7001T: < 0.1 ppm
GFC7001TM: <0.5 ppm
Span Drift (24 hours)
2
GFC7001T: < 0.5% of reading
GFC7001TM: 0.5ppm
Lag Time
1
10
seconds
Rise/Fall Time
1
<60 seconds to 95%
Linearity
GFC7001T: 1% of full scale
5
;
GFC7001TM: 0 - 3000 ppm: 1% full scale; 3000 - 5000 ppm: 2% full scale
Precision
GFC7001T: The greater of 0.5% of reading or 0.2ppm;
GFC7001TM: The greater of 1.0% of reading or 1ppm
Sample Flow Rate
800 cm
3
/min. ±10%
(O
2
Sensor option adds 120 cm³/min to total flow when installed)
AC Power
100V-120V, 220V-240V, 50/60 Hz
Analog Output Ranges
All Outputs: 10V, 5V, 1V, 0.1V (selectable)
Three outputs convertible to 4-20 mA isolated current loop.
All Ranges with 5% under/over-range
Analog Output Resolution
1 part in 4096 of selected full-scale voltage
Recorder Offset
±10%
Standard I/O
1 Ethernet: 10/100Base-T
2 RS-232 (300 – 115,200 baud)
2 USB device ports
8 opto-isolated digital status outputs
6 opto-isolated digital control inputs (2 defined, 4 spare)
4 user configurable analog outputs