9.6. CALIBRATION OF THE GFC7001T/GFC7001TM
ELECTRONIC SUBSYSTEMS
9.6.1. DARK CALIBRATION TEST
The dark calibration test interrupts the signal path between the IR photo-detector and the
remainder of the sync/demod board circuitry. This allows the instrument to compensate
for any voltage levels inherent in the sync/demod circuitry that might effect the
calculation of CO concentration.
Performing this calibration returns two offset voltages, one for CO MEAS and one for
CO REF that are automatically added to the CPU’s calculation routine. The two offset
voltages from the last calibration procedure may be reviewed by the user via the front
panel display. (See also Section 5.9.5).
To activate the dark calibration procedure or review the results of a previous calibration,
press:
Offset for CO REF signal
SAMPLE
RANGE=50.0 PPM CO= XX.XX
TST> CAL
SETUP
SETUP X.X
SECONDARY SETUP MENU
COMM VARS
DIAG
EXIT
SETUP X.X
PRIMARY SETUP MENU
CFG DAS RNGE PASS CLK
MORE
EXIT
DIAG
SIGNAL I/O
PREV
NEXT
ENTR
EXIT
SETUP X.X
ENTER PASSWORD
8
1
8
ENTR
EXIT
Continue pressing NEXT until ...
DIAG OPTIC
DARK CALIBRATION
PREV NEXT
ENTR
EXIT
DIAG DARK
CO DARK CALIBRATION
VIEW
CAL
EXIT
DIAG DARK
REF DARK OFFSET: 0.0mV
EXIT
Offset for CO MEAS signal
DIAG DARK
MEAS DARK OFFSET: 0.0mV
EXIT
DIAG DARK
DARK CAL 1% COMPLETE
EXIT
DIAG DARK
DARK CALIBRATION ABORTED
EXIT
Calibration runs automatically