beautypg.com

Dark calibration test, Teledyne analytical instruments 215 – Teledyne GFC-7001T - Trace CO Analyzer User Manual

Page 215

background image

Calibration Procedures Teledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

215

9.6. CALIBRATION OF THE GFC7001T/GFC7001TM

ELECTRONIC SUBSYSTEMS

9.6.1. DARK CALIBRATION TEST

The dark calibration test interrupts the signal path between the IR photo-detector and the
remainder of the sync/demod board circuitry. This allows the instrument to compensate
for any voltage levels inherent in the sync/demod circuitry that might effect the
calculation of CO concentration.
Performing this calibration returns two offset voltages, one for CO MEAS and one for
CO REF that are automatically added to the CPU’s calculation routine. The two offset
voltages from the last calibration procedure may be reviewed by the user via the front
panel display. (See also Section 5.9.5).
To activate the dark calibration procedure or review the results of a previous calibration,
press:

Offset for CO REF signal

SAMPLE

RANGE=50.0 PPM CO= XX.XX

TST> CAL

SETUP

SETUP X.X

SECONDARY SETUP MENU

COMM VARS

DIAG

EXIT

SETUP X.X

PRIMARY SETUP MENU

CFG DAS RNGE PASS CLK

MORE

EXIT

DIAG

SIGNAL I/O

PREV

NEXT

ENTR

EXIT

SETUP X.X

ENTER PASSWORD

8

1

8

ENTR

EXIT

Continue pressing NEXT until ...

DIAG OPTIC

DARK CALIBRATION

PREV NEXT

ENTR

EXIT

DIAG DARK

CO DARK CALIBRATION

VIEW

CAL

EXIT

DIAG DARK

REF DARK OFFSET: 0.0mV

EXIT

Offset for CO MEAS signal

DIAG DARK

MEAS DARK OFFSET: 0.0mV

EXIT

DIAG DARK

DARK CAL 1% COMPLETE

EXIT

DIAG DARK

DARK CALIBRATION ABORTED

EXIT

Calibration runs automatically