Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

73

Table 3-13 lists brief descriptions of the warning messages that may occur during start
up.

Table 3-13: Possible Warning Messages at Start-Up

Message MEANING

ANALOG CAL WARNING

The instrument's A/D circuitry or one of its analog outputs is not calibrated.

BENCH TEMP WARNING

Optical bench temperature is outside the specified limits.

BOX TEMP WARNING

The temperature inside the GFC7001T/GFC7001TM chassis is outside the
specified limits.

CANNOT DYN SPAN

2

Remote span calibration failed while the dynamic span feature was set to
turned on.

CANNOT DYN ZERO

3

Remote zero calibration failed while the dynamic zero feature was set to turned
on.

CONFIG INITIALIZED

Configuration was reset to factory defaults or was erased.

DATA INITIALIZED

DAS data storage was erased.

PHOTO TEMP WARNING

Photometer temperature outside of warning limits specified by
PHOTO_TEMP_SET variable.

REAR BOARD NOT DET

Motherboard was not detected during power up.

RELAY BOARD WARN

CPU is unable to communicate with the relay PCA.

SAMPLE FLOW WARN

The flow rate of the sample gas is outside the specified limits.

SAMPLE PRESS WARN

Sample pressure outside of operational parameters.

SAMPLE TEMP WARN

The temperature of the sample gas is outside the specified limits.

SOURCE WARNING

The IR source may be faulty.

SYSTEM RESET

1

The computer was rebooted.

WHEEL TEMP WARNING

The Gas Filter Correlation Wheel temperature is outside the specified limits.