14.4.2.5. PACKING COMPONENTS FOR RETURN TO TELEDYNE CUSTOMER

SERVICE

Always pack electronic components and assemblies to be sent to Teledyne Customer
Service in anti-ESD bins, tubes or bags.

CAUTION

ESD Hazard

DO NOT

use pink-poly bags.

NEVER

allow any standard plastic packaging materials to touch the

electronic component/assembly directly.

This includes

,

but is not limited to, plastic bubble-pack, Styrofoam

peanuts, open cell foam, closed cell foam, and adhesive tape.

DO NOT use standard adhesive tape as a sealer. Use ONLY anti-ESD tape.

Never carry the component or assembly without placing it in an anti-ESD bag or bin.
1. Before using the bag or container allow any surface charges on it to dissipate:

 If you are at the instrument rack, hold the bag in one hand while your wrist

strap is connected to a ground point.

 If you are at an anti-ESD workbench, lay the container down on the

conductive work surface.

 In either case wait several seconds.

2. Place the item in the container.
3. Seal the container. If using a bag, fold the end over and fastening it with anti-ESD

tape.

 Folding the open end over isolates the component(s) inside from the effects

of static fields.

 Leaving the bag open or simply stapling it shut without folding it closed

prevents the bag from forming a complete protective envelope around the
device.

Note

If you do not already have an adequate supply of anti-ESD bags or
containers available, Teledyne Customer Service department will supply
them (see Section 11.8 for contact information).

Follow the instructions listed above for working at the instrument rack
and workstation.