Getting StartedTeledyne API – Model T300/T300M CO Analyzer

Teledyne

Analytical

Instruments

58

Analyzer

Analyzer

Analyzer

Last Analyzer

Female DB9

Male DB9

RS-232

COM2

RS-232

COM2

RS-232

COM2

RS-232

COM2

Host

RS-232 port

Ensure jumper is

installed between

JP2 pins 21

22 in

last instrument of

multidrop chain.

Figure 3-17:

RS-232-Multidrop PCA Host

/

Analyzer Interconnect Diagram

7. BEFORE communicating from the host, power on the instruments and check

that the Machine ID code is unique for each (Section 5.7.1).
a. In the SETUP Mode menu go to SETUP>MORE>COMM>ID. The default

ID is typically the model number or “0”.

b. to change the identification number, press the button below the digit to

be changed.

c. Press/select ENTER to accept the new ID for that instrument.

8. Next, in the SETUP>MORE>COMM>COM1 menu (do not use the COM2

menu for multidrop), edit the COM1 MODE parameter as follows:
press/select EDIT and set only QUIET MODE, COMPUTER MODE, and
MULTIDROP MODE to ON. Do not change any other settings.

9. Press/select ENTER to accept the changed settings, and ensure that COM1

MODE now shows 35.

10. Press/select SET> to go to the COM1 BAUD RATE menu and ensure it

reads the same for all instruments (edit as needed so that all instruments are
set at the same baud rate).

Note

The Instrument ID’s should not be duplicated.

The (communication) Host instrument can only address one instrument
at a time.

Note

TAI recommends setting up the first link, between the Host and the first
analyzer, and testing it before setting up the rest of the chain.