CHAPTER 3

Installation Procedure

Before handling or unpacking MAGPIX, make certain that the selected site is appropriate.
See “Installation Diagrams” on page 28 for handling and site installation requirements and
detailed dimensions of MAGPIX.

Check for these requirements:

Indoors

Operating temperature of 15-35°C (59-95°F)

Operating relative humidity of 20-80%, noncondensing

Operating altitude up to 2400m (7874 ft) above mean sea level

Available electrical power outlet with protective earthing and easy accessibility

Available area of approximately 3‘ X 3’ (91.44cm), including a 2” (5.08cm) clearance
between the back of MAGPIX and any wall or vertical surface.

Stable, level surface

MAGPIX arrives in a large, corrugated cardboard overpack on skids.

FIGURE 9.

The MAGPIX Overpack

CAUTION:

This overpack is too heavy to be lifted by one person
(approximately 119lbs (53.97kg), a three-person lift) and should
be moved mechanically. Be careful the overpack is not punctured