calIbratIon Setup 3
Calibration #:
(LB)
c - u - l
The active calibration of four possible calibrations (16
calibrations possible with POS upgrade). The fields shown
on the Calibration Setup 1 screen represent the values of the
Calibration #
shown. Any edits made to the fields are
attributed to the Calibration # shown.
Linearization Mode:
(LB)
c
Disables and enables multi-point calibration.
Setup
disables multi-point calibration during normal deliveries.
Used when proving and setting multi-point calibrations.
Applied
enables multi-point calibration.
Linearization Prover Qty:
(fE)
c
The metered volume of the prover after a calibration delivery. When the prover volume is entered, the LCR 600 will
prompt you to select a linearization point number and a flow rate. After these values are entered, the LCR 600 will
assign the error accordingly. See calibration instructions.
BeGIN CALIBRATION deLIVeRY
Starts a calibration delivery.
Point #:
(fE)
c
The fields under Point #: are used to calibrate the meter/register at up to 10 different flow rates for each product
type/calibration.
Flow Rate:
(fE)
c
The flow rate at which the calibration point was measured.
Percent error:
(fE)
c
This field contains the percent difference between the register volume and the prover volume.
percent Error = [(prover Qty. - Meter Qty.) ÷ prover Qty.] x 100
calIbratIon Setup
l - locked
u - unlocked
c - Weights & measures calibration
p - Stop/pause
f - factory
Security clearance
fE - field Edit Box
LB - List Box