Calibration setup – Liquid Controls LCR 600 Setup & Op User Manual

Page 34

background image

34

calIbratIon Setup 3
Calibration #:

(LB)

c - u - l

The active calibration of four possible calibrations (16

calibrations possible with POS upgrade). The fields shown

on the Calibration Setup 1 screen represent the values of the
Calibration #

shown. Any edits made to the fields are

attributed to the Calibration # shown.

Linearization Mode:

(LB)

c

Disables and enables multi-point calibration.

Setup

disables multi-point calibration during normal deliveries.

Used when proving and setting multi-point calibrations.

Applied

enables multi-point calibration.

Linearization Prover Qty:

(fE)

c

The metered volume of the prover after a calibration delivery. When the prover volume is entered, the LCR 600 will

prompt you to select a linearization point number and a flow rate. After these values are entered, the LCR 600 will

assign the error accordingly. See calibration instructions.

BeGIN CALIBRATION deLIVeRY

Starts a calibration delivery.

Point #:

(fE)

c

The fields under Point #: are used to calibrate the meter/register at up to 10 different flow rates for each product

type/calibration.

Flow Rate:

(fE)

c

The flow rate at which the calibration point was measured.

Percent error:

(fE)

c

This field contains the percent difference between the register volume and the prover volume.

percent Error = [(prover Qty. - Meter Qty.) ÷ prover Qty.] x 100

calIbratIon Setup

l - locked
u - unlocked
c - Weights & measures calibration
p - Stop/pause
f - factory

Security clearance

fE - field Edit Box
LB - List Box

field type