Calibration setup, Compensation types and parameters, Single point or multi-point calibration – Liquid Controls LCR 600 Setup & Op User Manual

Page 28

background image

28

calIbratIon Setup

compensation types and parameters

product

vcf type

parameter

coefficient

range

ºcelsius/

ºfahrenheit tbase tmin

thold Tmax

General

Linear

Linear

0 to 0.003

ºC

15

-90

N/A

+100

General

Linear

Linear

0 to 0.005

ºF

60

-130

N/A

+212

LPG

USA

API Table 24

Specific Gravity

0.5 to 0.550

ºF

60

-50

-50

+140

LPG

Europe & Canada

API Table 54

Density

kg/L

0.5 to 0.600

ºC

15

-46

-46

+60

Refined Petro-
leum Products

Europe & Canada

API Table 54B

Density

kg/m3

653.0 to 1075.0

ºC

15

-50

-40

+95

Refined Petro-
leum Products

USA

API Table 6B

API Gravity

0 to 85

ºF

60

-50

-40

+200

General

API Table 54C

Coefficient

0.000486 to

0.001674

ºC

15

-50

-40

+95

Lube Oil

Europe & Canada

API Table 54B

Density

kg/m3

800 to 1164

ºC

15

-50

-40

+95

Ammonia

Canada

NH3

N/A

N/A

ºC

15

-30

-30

+40

Single point or multi-point calibration

The LCR 600 provides two means of calibration, single point or multi-point calibration. A “point” corresponds to a

particular flow rate along the meter’s accuracy curve. Single point calibration adjusts one point along the accuracy

curve to zero percent error—typically at a flow rate representative of a normal delivery. Multi-point calibration zeroes

the percent error at multiple flow rates in order to flatten out the accuracy curve across the meter’s range of flow rates.

Multi-point calibration is ideal for applications in which deliveries are made at many different flow rates, while single

point calibration is appropriate for applications where a high flow rate is typical for most deliveries.

Single Point Calibration Accuracy Curve

Multi-point Calibration Accuracy Curve

Single point and multi-point calibration accuracy curves