calIbratIon Setup
compensation types and parameters
product
vcf type
parameter
coefficient
range
ºcelsius/
ºfahrenheit tbase tmin
thold Tmax
General
Linear
Linear
0 to 0.003
ºC
15
-90
N/A
+100
General
Linear
Linear
0 to 0.005
ºF
60
-130
N/A
+212
LPG
USA
API Table 24
Specific Gravity
0.5 to 0.550
ºF
60
-50
-50
+140
LPG
Europe & Canada
API Table 54
Density
kg/L
0.5 to 0.600
ºC
15
-46
-46
+60
Refined Petro-
leum Products
Europe & Canada
API Table 54B
Density
kg/m3
653.0 to 1075.0
ºC
15
-50
-40
+95
Refined Petro-
leum Products
USA
API Table 6B
API Gravity
0 to 85
ºF
60
-50
-40
+200
General
API Table 54C
Coefficient
0.000486 to
0.001674
ºC
15
-50
-40
+95
Lube Oil
Europe & Canada
API Table 54B
Density
kg/m3
800 to 1164
ºC
15
-50
-40
+95
Ammonia
Canada
NH3
N/A
N/A
ºC
15
-30
-30
+40
Single point or multi-point calibration
The LCR 600 provides two means of calibration, single point or multi-point calibration. A “point” corresponds to a
particular flow rate along the meter’s accuracy curve. Single point calibration adjusts one point along the accuracy
curve to zero percent error—typically at a flow rate representative of a normal delivery. Multi-point calibration zeroes
the percent error at multiple flow rates in order to flatten out the accuracy curve across the meter’s range of flow rates.
Multi-point calibration is ideal for applications in which deliveries are made at many different flow rates, while single
point calibration is appropriate for applications where a high flow rate is typical for most deliveries.
Single Point Calibration Accuracy Curve
Multi-point Calibration Accuracy Curve
Single point and multi-point calibration accuracy curves