calIbratIon Setup

multi-point calibration proving - continued

LCR-600: 60011

Adjacent linearization points out of

0.25% range.

The LCR 600 will not allow you to enter the Applied

setting of the Linearization Mode: until all adjacent

calibration points are within 0.25% of each other.

to prove a meter system to a product with multi-

point calibration:

10. Repeat steps 4 through 10 using a new flow rate for the

next Point #: until all the points between the lowest

flow rate and the maximum flow rate are within 0.25% of
each other.

11. Navigate to Calibration Setup 1. Move the pointer

 to

Linearization Mode:

and select Applied.

The linearization points on Calibration

Setup 3 will automatically sort by flow rate

after the Applied

setting is selected.