Burkert Type 8076 User Manual
Page 61

59
Adjustmentandcommissioning
Detailed procedure of a teach-in by flow rate (“flow.teach” function in the “sensor” menu)
→
Choose the volume unit in which the teach-in is run:
Calib
Sensor
Flow_L
Flow_m3
Flow_gal
Flow_Igal
PVar:
Flow Teach
Freq.
→
Run the teach-in:
→
Establish the flow in the pipe and wait until it is
stabilised.
→
Save the new K
factor calculated or
not.
Flow Teach
Sensor
Start Teach
Calib
0005 L/min
Measure
K teached:
K=3,810
Cal. Result
BACK
Save: Yes/No
Cal. Result
The device displays the K factor
calculated by teach-in.
OK
The device calculates the flow measured
in the pipe, with the current K factor, for
100 seconds
1)
.
The bargraph shows the time passed.
FlowTeach
0049 L/min
0001 L/s
The device displays the average flow rate
calculated with the current K factor.
→
Enter the value of the flow in
the pipe.
→
Modify the time unit if
necessary.
OK
K teached set to
max K=9999
BACK
→
Confirm the K factor =9999 with “YES”
or keep the current K factor with “NO”.
Save modified
data? Yes/No
If the calculated K factor
is ≥ 9999, it is forced to
9999.
1)
Measurement can be suspended at any time by selecting “OK”.
English
Type 8026- 8036- 8076