Adjustmentandcommissioning
→
Run the teach-in:
→
Open the valve to fill the tank.
→
Save the new K
factor calculated or
not.
Vol.Teach
Sensor
Start Teach
Calib
K teached:
K=3,810
Cal. Result
BACK
Save: Yes/No
Cal. Result
The device displays the K factor
calculated by teach-in.
OK
The device displays the instantaneous
flow of the fluid.
OK
K teached set to
max K=9999
BACK
→
Confirm the K factor =9999 with “YES”
or keep the current K factor with “NO”.
If the calculated K factor is
≥ 9999, it is forced to 9999.
Meas. Progr.
→
When the tank is filled, close the valve.
OK
2.001 l/s
The device displays the volume calcu-
lated with the current K factor. If the
display shows “++++”, the volume
calculated > 9999 volume units: conti-
nue the procedure as normal.
→
Enter the actual volume of fluid
in the tank.
Input Vol.
101.2 l
+099.0 l
Save modified
data? Yes/No