See chap. 9.10 to access the calib menu, English – Burkert Type 8076 User Manual
Page 59
57
Adjustmentandcommissioning
9.13.5. entering the K factor or determining it with Teach-in
See chap. 9.10 to access the Calib menu.
K.fact.
Calib
Sensor
INPUT
Flow_L
Flow_m3
Flow_gal
Flow_Igal
PVar:
Start Teach
Vol.Teach
Flow Teach
Measure
Flow Teach
Cal. Result
Litre
m3
gal
Igal
PVar:
Start Teach
Meas.Progress
Input Vol.
Cal. Result
Freq.
K.FACT.: Enter the K factor of the fitting used.
VOL.TEACH: Determine the fitting K factor using a teach-in procedure depending on a volume. See details of the
procedure below.
FLOW TEACH: Determine the fitting K factor using a teach-in procedure depending on the flow rate. See details
of the procedure on page 59
Detailed procedure of a teach-in by volume (“Vol.teach” function in the “sensor” menu)
→
Prepare a tank capable of containing 100 litres, for example.
→
Choose the volume unit in which the teach-in is run:
Calib
Sensor
L
m3
gal
Igal
PVar:
Vol.Teach