beautypg.com

See chap. 9.10 to access the calib menu, English – Burkert Type 8076 User Manual

Page 59

background image

57

Adjustmentandcommissioning

9.13.5. entering the K factor or determining it with Teach-in

See chap. 9.10 to access the Calib menu.

K.fact.

Calib

Sensor

INPUT

Flow_L

Flow_m3

Flow_gal

Flow_Igal

PVar:

Start Teach

Vol.Teach

Flow Teach

Measure

Flow Teach

Cal. Result

Litre

m3

gal

Igal

PVar:

Start Teach

Meas.Progress

Input Vol.

Cal. Result

Freq.

K.FACT.: Enter the K factor of the fitting used.
VOL.TEACH: Determine the fitting K factor using a teach-in procedure depending on a volume. See details of the
procedure below.
FLOW TEACH: Determine the fitting K factor using a teach-in procedure depending on the flow rate. See details
of the procedure on page 59

Detailed procedure of a teach-in by volume (“Vol.teach” function in the “sensor” menu)

Prepare a tank capable of containing 100 litres, for example.

Choose the volume unit in which the teach-in is run:

Calib

Sensor

L

m3

gal

Igal

PVar:

Vol.Teach

English

Type 8026- 8036- 8076

This manual is related to the following products: