Modifying the param menu access code, See chap. 9.10 to access the param menu, F pvar = freq. if pvar ≠ freq – Burkert Type 8076 User Manual
Page 50
48
Adjustmentandcommissioning
9.12.2. modifying the pArAm menu access code
See chap. 9.10 to access the Param menu. The default access code is "0000".
Code
0***
Confirm
0***
System
Param
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
Enter the new
access code to
the PARAM menu
Confirm the new
code
9.12.3. restoring the default parameters of the process level and
the outputs
See chap. 9.10 to access the Param menu.
Factory Set
Execute
Reset Yes/No
System
Param
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
EXECUTE : restore (choose "Yes") the default parameters of the Process level and of the outputs, or keep the
current parameters ( choose "No").
9.12.4. setting the data displayed on lines 1 and 2 of the display
module
See chap. 9.10 to access the Param menu.
This is
when the
device is be-
ing parame-
tered............
....................
This is
when the
device is be-
ing parame-
tered............
....................
Display
Param
Line1 / Line2
PVar:
Unit:
Filter:
Flow_L
Flow_m3
Flow_gal
Flow_Igal
/h
Slow
Fast
None
Freq.
/min
/s
Hz
f PVar = Freq.
If PVar ≠ Freq.
Enabled
Disabled
Line 1/Line2:
Activate or deactivate line 1 or line 2 respectively in process level.
LINE1 or LINE2 : Activate (choice “Enabled”) or deactivate (choice “Disabled”) the display of line 1 or line 2
respectively.
set the parameters of the data displayed in process level on line 1 or line 2 respectively when the corres-
ponding line is activated:
PVAR: Choose the flow rate volume unit or the input frequency to be displayed on line 1 or line 2 of the display
respectively.