:AUTO

Command Syntax:

:TRIGger:DELay:AUTO

Command parameter:

= 1 or ON

Enable auto delay

0 or OFF

Disable auto delay

Query:

:AUTO?

Query delay state

Return String:

1 or 0

Description:

This command is used for enable or disable trigger auto delay. If disabled, the trigger will delay

based

on the delay time configured with “:TRIGger:DELay” command.

6.3.8 Common Commands

Common commands can be used in instruments that are SCPI-compliant. The common commands

available for the 5492B are described below:

*RST

Command Syntax:

*RST

Description:

Resets the instrument.

*TRG

Command Syntax:

*TRG

Description:

Sends a bug trigger to the instrument. Use this if trigger source is set to BUS.

*IDN?

Query Syntax:

*IDN?

Query return:

,,

Example: 5492B Digital Multimeter, Ver1.0.00.00.01,123A45678

Description: Query the identification of the instrument.