SCPI Command Reference

98

of all programmed operations, the instrument will return to the top of the trigger model.

NOTE: With continuous initiation enabled (ON), you cannot use the :READ? Command.

:ABORt

Command Syntax:

:ABORt

Description:

When this command is sent, the multimeter aborts operation and returns to the top of the Trigger

Model.

:TRIGger

This subsystem is used for to configure trigger mode and trigger delay

:SOURce

Command Syntax:

TRIGger:SOURce

Command parameter:

= IMMediate

Internal trigger

BUS Trigger via USB/RS-232 or GPIB interface

MANual

Press

in the front panel for Trigger

EXTernal

Trigger from external signal connected to EXT TRIG terminal

Query:

:SOURce?

Query trigger source

Return String:

IMM, BUS, MAN, or EXT

Description:

This command is used to select the trigger source.

:DELay

Command Syntax:

:TRIGger:DELay

Command parameter:

= 0 to 6000

Specify delay in milliseconds

Description:

This command is used to set the delay time for trigger delay. It takes into effect only if auto delay is

disabled.