SCPI Command Reference

90

:COUNt

Command syntax:

[:SENSe[1]]:CURRent:AC:AVERage:COUNt

[:SENSe[1]]:CURRent:DC:AVERage:COUNt

[:SENSe[1]]:VOLTage:AC:AVERage:COUNt

[:SENSe[1]]:VOLTage:DC:AVERage:COUNt

[:SENSe[1]]:RESistance:AVERage:COUNt

[:SENSe[1]]:FRESistance:AVERage:COUNt

Command parameter:

= 1 to 100

Specify filter count

MIMimum 1

MAXimum 100

Query:

:COUNt?

Query filter count

Description:

These commands are used to specify the filter count. In general, the filter count is the number of

readings that are acquired and stored in the filter buffer for the averaging calculation. The larger the

filter count, the more filtering that is performed.

THReshold Command

Use these commands to set the maximum range input (signal level) for frequency and period

measurements.

:RANGe

Command syntax:

[:SENSe[1]]:FREQuency:THReshold:VOLTage:RANGe Set voltage threshold range.

[:SENSe[1]]:PERiod:THReshold:VOLTage:RANGe

Set voltage threshold range.

Command parameter:

= 0.1 (120 mV), 1 (1.2 V), 10 (12 V), 100 (120 V), 750 (750 V)

Specify signal levels in volts (voltage threshold)

Query:

:RANGe?

Query maximum signal level.

Return String:

0.1, 1, 10, 100, or 750

Description:

This command is used to set the current or voltage threshold range for frequency and period