SCPI Command Reference
78
Description:
This command is used to specify the upper and lower limit for limit testing. The actual limit depends
on which measurement function is currently selected. For example, a limit value of 1 is 1V for the
voltage measurement functions (DCV or ACV), 1A for the current measurement functions (DCI or
ACI), 1
Ω on the resistance measurement function (2-wire or 4-wire). A limit value is not range
sensitive. For example, a limit of 1 for DCV is 1 V on all measurement ranges.
:STATe
Command syntax:
:CALCulate3:LIMit[1]:STATe
Command Parameter:
= 0 or OFF
Disable limit test
1 or ON
Enable limit test
Query:
:STATe?
Query state (on or off) of limit test
Return String:
1 or 0
Description:
This command is used to enable or disable the limit test function. When enabled, the limit test will be
active.
:FAIL?
Command syntax:
:CALCulate3:LIMit[1]:FAIL?
Description:
This command is used to read the results of the limit test
0 = Limit test failed
1 = Limit test passed
The response message (0 or 1) only tells you if a limit test has passed or failed. In the case of 0
(failed), it does not tell you which limit (upper or lower) has failed.