B&K Precision 5492BGPIB - Manual User Manual

Page 78

background image

SCPI Command Reference

78

Description:

This command is used to specify the upper and lower limit for limit testing. The actual limit depends

on which measurement function is currently selected. For example, a limit value of 1 is 1V for the

voltage measurement functions (DCV or ACV), 1A for the current measurement functions (DCI or

ACI), 1

Ω on the resistance measurement function (2-wire or 4-wire). A limit value is not range

sensitive. For example, a limit of 1 for DCV is 1 V on all measurement ranges.

:STATe

Command syntax:

:CALCulate3:LIMit[1]:STATe

Command Parameter:

= 0 or OFF

Disable limit test

1 or ON

Enable limit test

Query:

:STATe?

Query state (on or off) of limit test

Return String:

1 or 0

Description:

This command is used to enable or disable the limit test function. When enabled, the limit test will be

active.

:FAIL?

Command syntax:

:CALCulate3:LIMit[1]:FAIL?

Description:

This command is used to read the results of the limit test

0 = Limit test failed

1 = Limit test passed

The response message (0 or 1) only tells you if a limit test has passed or failed. In the case of 0

(failed), it does not tell you which limit (upper or lower) has failed.