SCPI Command Reference

74

Description:

This command is used to enable or disable the CALC1 calculation. When enabled, each instrument

reading will reflect the selected calculation.

:DATA?

Command syntax:

:CALCulate[1]:DATA?

Return String:

Math calculated measurement in format

Description:

This query command is used to read the result of the CALC1 calculation. If CALC1 is disabled or

NONE is selected, the measured reading without the math calculations will be returned.

:CALCulate2 Subsystem

Use this subsystem for commands to configure and control buffer operations.

:TRACe Command

:CLEar

Command syntax:

:CALCulate2:TRACe:CLEar

Description:

This action command is used to clear the readings in the buffer. If the buffer is not cleared, a

subsequent store will overwrite the previous readings. If the subsequent store is aborted before the

buffer becomes full, you could end up with some readings that were previously stored in the buffer.

Note: For USB(virtual com) and RS-232 interface, this command must be used before a

subsequent store can

process. You will know that the display will not have “*” lid up until the

buffer is cleared, even if store reading function is enabled with “:CALC2:STAT 1” command.

:POINTS

Command syntax:

:CALCulate2:TRACe:POINTS

Command Parameter:

=

2 to 512 Specify buffer size.

Query:

:POINTS?

Query buffer size.

Description:

This command is used to specify the buffer size, which in turn sets the number of readings to store.