B&K Precision 5492BGPIB - Manual User Manual
Page 74

SCPI Command Reference
74
Description:
This command is used to enable or disable the CALC1 calculation. When enabled, each instrument
reading will reflect the selected calculation.
:DATA?
Command syntax:
:CALCulate[1]:DATA?
Return String:
Math calculated measurement in
Description:
This query command is used to read the result of the CALC1 calculation. If CALC1 is disabled or
NONE is selected, the measured reading without the math calculations will be returned.
:CALCulate2 Subsystem
Use this subsystem for commands to configure and control buffer operations.
:TRACe Command
:CLEar
Command syntax:
:CALCulate2:TRACe:CLEar
Description:
This action command is used to clear the readings in the buffer. If the buffer is not cleared, a
subsequent store will overwrite the previous readings. If the subsequent store is aborted before the
buffer becomes full, you could end up with some readings that were previously stored in the buffer.
Note: For USB(virtual com) and RS-232 interface, this command must be used before a
subsequent store can
process. You will know that the display will not have “*” lid up until the
buffer is cleared, even if store reading function is enabled with “:CALC2:STAT 1” command.
:POINTS
Command syntax:
:CALCulate2:TRACe:POINTS
Command Parameter:
2 to 512 Specify buffer size.
Query:
:POINTS?
Query buffer size.
Description:
This command is used to specify the buffer size, which in turn sets the number of readings to store.