2 front panel fuse replacement, Basic measurements, Current source – B&K Precision 5492BGPIB - Manual User Manual
Page 25

Basic Measurements
25
CATⅡ(300V)
CATⅠ(1000V)
1000V
MAX
Current Source
▲
▲
▲
▲
DC V
AC V
2W
Fr eq
Aut o
Tr i
g
MX+B
Shi f t
Cont
Rel
¦ ¸
Period
dB/m
Fast
Menu
Recall
Med
Slow
Hold
CHOICES
LEVEL
ENTER
ESC
LOCAL
%
I
DC
I
AC
4W
Ω
FAST
MED
SLOW
ADRS
RMT
HOLD TRIG
*
MEM
AUTO
REL
FILT MATH
SHIFT
4W
ERR
POWER
T2AL 250V
SENSE
Ω
4W
VΩ
HI
!
10A
CATⅡ(300V)
350V
MAX
12A
MAX
mA
12A
INPUT
R
5
5492B
1 2
/
Digit Multimeter
For low current measurement: INPUT LO and SENSE LO terminals are used.
CAUTION: Maximum input = 1 A DC or RMS
1000V
MAX
12A
MAX
12A
Current Source
▲
▲
▲
▲
DC V
AC V
2W
Fr eq
Aut o
Tr i g
MX+B
Shi f t
Cont
Rel
¦ ¸
Period
dB/m
Fast
Menu
Recall
Med
Slow
Hold
CHOICES
LEVEL
ENTER
ESC
LOCAL
%
I
DC
I
AC
4W
Ω
FAST
MED
SLOW
ADRS
RMT
HOLD TRIG
*
MEM
AUTO
REL
FILT MATH
SHIFT
4W
ERR
POWER
T2AL 250V
SENSE
Ω
4W
VΩ
LO
HI
!
10A
CATⅡ(300V)
CATⅠ(1000V)
350V
MAX
mA
INPUT
R
5
5492B
1 2
/
Digit Multimeter
For 1 A or higher current measurement: 10 A and INPUT LO terminals are used.
CAUTION: Maximum Input = 12 A DC or RMS
Note: Auto range is not available for 1.2 A and 12 A ranges
Figure 3-2
DC and AC Current Measurements
3.3.2 Front Panel Fuse Replacement
WARNING: Make sure the instrument is disconnected from the power line and other equipment
before replacing the fuse.
1. Turn off the power and disconnect the power line and test leads.
2. From the front panel, use a screwdriver to rotate the fuse holder several turns counter-clockwise.
Take the fuse carrier out of the socket.
3. Remove the fuse and replace it with the same type (T2AL, 250 V, 5×20mm)
CAUTION: Do not use a fuse with a higher current rating than specified or instrument damage
may occur. If the instrument repeatedly blows fuses, try to find out the reason
before replacing the fuse.