Page 76

SCPI Command Reference
76
:FORMat
Command syntax:
:CALCulate2:FORMat
Command Parameter:
No calculation
MAXimum Largest reading in buffer
MINimum Lowest reading in buffer
MEAN
Mean (average) value of readings in buffer
SEDViation Standard deviation of readings in buffer
Query:
:FORMat?
Query programmed math format.
Return String:
NONE, MEAN, SEDV, MAX, or MIN
Description:
This command is used to specify
which statistical calculations to recall (using “:CALC2:DATA?”
command) from the stored readings in the buffer. If NONE is selected or if the buffer is empty,
using “:CALC2:DATA?” command will return the current measured reading on the display.
:STATe
Command syntax:
:CALCulate2:STATe
Command Parameter:
= 0 or OFF
Disable store reading
1 or ON
Enable store reading
Query:
:STATe?
Query state (ON or OFF) of store reading function
Return String:
1 or 0
Description:
This command is used to enable or disable
store reading. It is equivalent to setting “RDGS
STORE” to ON from the front panel menu. When enabled, readings on the display will immediately
begin storing into the buffer if the meter is already in measurement mode.
For USB/RS-232 Interface: To store reading subsequently with this command, the buffer must be
cleared prior to sending this command using the “:CALC2:TRAC:CLE” command.