83
Table 6-4
SENSe Command Summary (cont.)
Command
Function description
Default
:DIODe
:CURRent
:RANGe
[:UPPer]
[:UPPer]?
Path to configure diode test
Path to select range
Select range
Query range
1e-3
:CONTinuity
:THReshold
:THReshold?
Path to configure continuity test
Specify threshold resistance (1 to 1000)
Query threshold resistance
10
:[SENSe[1]] Subsystem
:FUNCtion
Command syntax:
[:SENSe[1]]:FUNCtion
Command Parameter:
VOLTage:AC
Select AC Voltage
VOLTage:DC Select DC Voltage
CURRent:AC Select AC Current
CURRent:DC
Select DC Current
RESistance
Select 2-wire Resistance
FRESistance
Select 4-wire Resistance
FREQuency Select Frequency
PERiod Select Period
DIODe Select Diode Testing
CONTinuity
Select Continuity Testing
Query:
:FUNCtion?
Query currently programmed function
Description:
This command is used to select the measurement function of the instrument.
Each measurement function “remembers” its own unique setup configuration, such as range and
speed. This eliminates the need to re-program setup conditions every time you switch from one
function to another.