SCPI Command Reference
68
CONFigure Command
Command syntax:
:CONFigure:
Command parameter :
VOLTage:AC AC voltage
CURRent:DC
DC current
CURRent:AC AC current
RESistance 2-wire resistor
FRESistance 4-wire resistor
FREQuency Frequency
PERiod Period
DIODe Diode testing
CONTinuity Continuity test
Query:
:CONFigure?
Query the selected function.
Return String Example:
volt:dc
Description:
This command configures the instrument for subsequent measurements on the specified function. It
places the instrument in the selected measurement mode. You then use the :READ? command to
trigger a measurement and acquire a reading.
When this command is sent, the multimeter will be configured as follows:
The function specified by this command is selected.
All controls related to the selected function are defaulted to the *RST values.
Continuous initiation is disabled.
The control source of the Trigger Model is set to Immediate.
The delay of the Trigger Model is set to zero.
All math calculations are disabled.
Buffer operation is disabled. A storage operation currently in process will be aborted.
Autozero is set to the *RST default value.
This command is automatically asserted when the :MEASure? Command is sent.