SCPI Command Reference

68

CONFigure Command

Command syntax:

:CONFigure:

Command parameter :

= VOLTage:DC DC voltage

VOLTage:AC AC voltage

CURRent:DC

DC current

CURRent:AC AC current

RESistance 2-wire resistor

FRESistance 4-wire resistor

FREQuency Frequency

PERiod Period

DIODe Diode testing

CONTinuity Continuity test

Query:

:CONFigure?

Query the selected function.

Return String Example:

volt:dc

Description:

This command configures the instrument for subsequent measurements on the specified function. It

places the instrument in the selected measurement mode. You then use the :READ? command to

trigger a measurement and acquire a reading.

When this command is sent, the multimeter will be configured as follows:
 The function specified by this command is selected.
 All controls related to the selected function are defaulted to the *RST values.
 Continuous initiation is disabled.
 The control source of the Trigger Model is set to Immediate.
 The delay of the Trigger Model is set to zero.
 All math calculations are disabled.
 Buffer operation is disabled. A storage operation currently in process will be aborted.
 Autozero is set to the *RST default value.
This command is automatically asserted when the :MEASure? Command is sent.