SCPI Command Reference

97

6.3.7 TRIGger Subsystem

The Trigger subsystem is made up of a series of commands to configure the Trigger model. These

commands and subsystems are summarized in Table 6-7

Table 6-7

TRIGger Command Summary

Command

Function Description

Default

:INITiate

[:IMMediate]

:CONTinuous

:CONTinuous?

:ABORt

:TRIGger

:SOURce

:SOURce?

:DELay

:AUTO

:AUTO?

Subsystem command path

Initiate one trigger cycle.

Enable or disable continuous initiation.

Query continuous initiation.

Reset trigger system

Select control source

Query control source

Set delay time

Enable or disable auto delay

Query state of delay

IMMediate

0

OFF

:INITiate

[:IMMediate]

Command Syntax:

:INITiate[:IMMediate]

Description:

This command takes the instrument out of the idle state. After all programmed operations are

completed, the instrument returns to the idle state if continuous initiation is disabled (see next

command).

:CONTinuous

Command Syntax:

:INITiate:CONTinuous

Control continuous initiation

Command parameter:

= 0 or OFF

Disable continuous initiation

1 or ON

Enable continuous initiation

Query:

:CONTinuous?

Query continuous initiation

Return String:

1 or 0

Description:

When continuous initiation is selected (ON), the instrument is taken out of the idle state. At the end