beautypg.com

2 display subsystem, 3 calculate subsystem – B&K Precision 5492BGPIB - Manual User Manual

Page 70

background image

SCPI Command Reference

70

6.3.2 DISPlay subsystem

The DISPlay subsystem commands are mainly used to control the display of the multimeter and are

summarized in Table 6-2.

Table 6-2

DISPlay Subsystem Commands Summary

Command

Function Description

:DISPlay

:ENABle

:ENABle?

Enable or disable front panel display

Query state of the display

:ENABle

Command syntax:

:DISPlay:ENABle

Command Parameter:

=

0 or OFF

Disable front panel display

1 or ON

Enable front panel display

Query:

:ENABle? Query state of the display

Return String:

1 or 0

Description:

This command is used to enable or disable front panel display circuitry. When disabled, the

instrument operates at a higher speed. While disabled, the display will be frozen. All front panel

controls except LOCAL are disabled. Normal display operation can be resumed by using :ENABle

command or pressing LOCAL key to enable the display.

6.3.3 CALCulate Subsystem

The commands in this subsystem are used to configure and control the calculate subsystems and are

summarized in Table 6-3.

Table 6-3

CALCulate Command Summary

Command

Function Description

Default

Parameter

:CALCulate[1]

:FORMat

:FORMat?

:KMATh

:MMFactor

:MMFactor?

Subsystem to control CALC1 (MX+B and % MATH)

Select math format (NONE,MXB,PERCent)

Query math format.

Path to configure math calculations:
Set “m” factor for mx+b (-100e6 to 100e6)
Q

uery “m” factor

NONE

1

This manual is related to the following products: