SCPI Command Reference

96

Description:

The command is used to select the units for DCV measurement. With volt (V) unit selected, normal

AC voltage measurement are made. With DB unit selected, DC dB voltage measurements are

performed. With DBM units selected, DC dBm voltage measurements are made.

:DB:REFerence

Command Syntax:

:UNIT:VOLTage:DC:DB:REFerence

Command parameter:

= 1e-7 to 1000

Specify reference voltage in volts

Query:

:DB:REFerence?

Query reference voltage

Description:

This command is used to specify the dB reference level. When DB unit is selected, DCV dB

measurements are made using the specified reference voltage.

The reference level is specified in volts and is not range dependent. For example: a dB reference

level of 1 is 1 V on all DCV measurement ranges.

:DBM:IMPedance

Command Syntax:

:UNIT:VOLTage:DC:DBM:IMPedance

Command Parameter:

= 1 to 9999

Specify reference impedance

Query:

:DBM:IMPedance? Query reference impedance

Description:

This command is used to specify the dBm reference impedance level. When dBm unit is selected,

DCV dBm measurements are made using the specified reference impedance.

The reference impedance is specified in ohms and is not range dependent. For Example: a dBm

reference level of 500 is 500 on all DCV measurement ranges.