B&K Precision 5492BGPIB - Manual User Manual
Page 96

SCPI Command Reference
96
Description:
The command is used to select the units for DCV measurement. With volt (V) unit selected, normal
AC voltage measurement are made. With DB unit selected, DC dB voltage measurements are
performed. With DBM units selected, DC dBm voltage measurements are made.
:DB:REFerence
Command Syntax:
:UNIT:VOLTage:DC:DB:REFerence
Command parameter:
Specify reference voltage in volts
Query:
:DB:REFerence?
Query reference voltage
Description:
This command is used to specify the dB reference level. When DB unit is selected, DCV dB
measurements are made using the specified reference voltage.
The reference level is specified in volts and is not range dependent. For example: a dB reference
level of 1 is 1 V on all DCV measurement ranges.
:DBM:IMPedance
Command Syntax:
:UNIT:VOLTage:DC:DBM:IMPedance
Command Parameter:
Specify reference impedance
Query:
:DBM:IMPedance? Query reference impedance
Description:
This command is used to specify the dBm reference impedance level. When dBm unit is selected,
DCV dBm measurements are made using the specified reference impedance.
The reference impedance is specified in ohms and is not range dependent. For Example: a dBm
reference level of 500 is 500 on all DCV measurement ranges.