B&K Precision 4045B - Manual User Manual
Page 87

87
AMPL 2.5V
AMPL MAX
Query
Syntax: AMPLitude?
Examples: AMPL?
AMPL? MAX
Response: NR2
Considerations:
1) The MAXimum amplitude is dependent on the offset.
2) The MAX and MIN arguments should not be used in a
program message containing an OFFSet command, since
these values are evaluated during parsing, based on the
current value of the offset.
OFFSet
The offset command is used to set the DC offset of the
output waveform. Note that the amplitude and the offset
are limited by the relation:
Peak Amplitude + |Offset| ≤ 5V
Arguments
Type: Numeric
Units: V, mV
Range: 10mV to 5V
Rounding: To 10mV
Command Type: Setting or Query
Setting
Syntax: OFFSet
See also other documents in the category B&K Precision Generators:
- 4040B - Datasheet (3 pages)
- 4040A - Datasheet (1 page)
- 4030 - Datasheet (1 page)
- 4012A - Datasheet (1 page)
- 4013B - Datasheet (2 pages)
- 4005DDS - Datasheet (1 page)
- 4011A - Datasheet (1 page)
- 3003 - Datasheet (1 page)
- 4003A - Datasheet (1 page)
- 2005B - Datasheet (1 page)
- 4001A - Manual (19 pages)
- 4010A - Manual (39 pages)
- 4005DDS - Manual (21 pages)
- 4011A - Manual (39 pages)
- 4007B - Manual (20 pages)
- 4030 - Manual (21 pages)
- 4017A - Manual (21 pages)
- 4014B - Manual (84 pages)
- 4052-4055 - Manual (78 pages)
- 4052-4055 - Datasheet (5 pages)
- 4052-4055 - Programming Manual (33 pages)
- 4040A - Manual (25 pages)
- 4045B - Datasheet (3 pages)
- 4084-4087 - Manual (60 pages)
- 4084-4087 - Datasheet (2 pages)
- 4075 - Manual (103 pages)
- 4075 - Datasheet (4 pages)
- 4084AWG - Datasheet (2 pages)
- 4076 - Manual (106 pages)
- 4076 - Datasheet (4 pages)
- 4033 - Manual (72 pages)
- 4033 - Datasheet (3 pages)
- 262 - Datasheet (1 page)
- 1275 - Datasheet (1 page)
- 1211E - Datasheet (1 page)
- 1280B - Manual (8 pages)
- 1280B - Datasheet (1 page)
- 1249B - Manual (41 pages)
- 1253 - Datasheet (1 page)
- 1253 - Manual (37 pages)