19
Verify that all items above are included in the shipping
container. If anything is missing, please contact B&K
Precision.
2.3 Instrument Mounting
The 4045B Arbitrary Function Generator is intended for
bench use. The instrument includes a front feet tilt
mechanism for optimum panel viewing angle. The
instrument does not require special cooling when
operated within conventional temperature limits. It may
be installed in a closed rack or test station if proper air
flow can assure removing about 15 W of power
dissipation.
2.4 Power Requirements
The 4045B can be operated from any source of 90V to
264V AC, frequency from 48Hz to 66Hz. The maximum
power consumption is 30VA. Use a slow blow fuse of 1A,
UL/CSA approved as indicated on the rear panel of the
instrument.
The instrument power fuse is located in the AC input
plug. To access the fuse, first disconnect the power
cord and then remove the fuse box.