B&K Precision 4045B - Manual User Manual

Page 26

background image

26

Excessive distortion or aberrations caused by

improper termination are less noticeable at lower

frequencies, especially with sine and triangle

waveforms. To ensure waveform integrity, follow

these precautions:

1. Use good quality 50 Ω coaxial cable and

connectors.

2. Make all connections tight and as short as

possible.

3. Use good quality attenuators if it is necessary to

reduce waveform amplitudes applied to sensitive

circuits.

4. Use termination or impedance-matching devices

to avoid reflections.

5. Ensure that attenuators and terminations have

adequate power handling capabilities.

If there is a DC voltage across the output load, use a

coupling capacitor in series with the load. The time

constant of the coupling capacitor and load must be

long enough to maintain pulse flatness.

Impedance Matching

If the waveform generator is driving a high

impedance, such as a 1 MΩ input impedance

(paralleled by a stated capacitance) of an

oscilloscope vertical input, connect the transmission

line to a 50 Ω attenuator, a 50 Ω termination and to

the oscilloscope input. The attenuator isolates the

input capacitance of the device and terminates the

waveform generator properly.